User Manual 301/302
When the main output state is programmed to ON, the output
connector is connected to the output amplifier through a 50Ω resistor
and an LED near the output connector illuminates. In the OFF
position, the output connector is disconnected from the output
amplifier by means of a mechanical relay. Ensure that voltage is not
applied to the main output connector when the Model 301/302 output
state is programmed to ON.
The query:
OUTPut?
queries the state of the main output and returns "0" (OFF) or "1" (ON).
Figure 3.6 Below shows how the Instrument control panel in WaveWiz
can be used to enable the main outputs. Checking the relevant box
will toggle each output on or off.
Selecting an
Output Type
There are three main types of waveforms that the model 301/302 can
produce: Standard (fixed), Arbitrary (user) and Sequenced waveforms.
Standard waveforms are computed from formulas that are built into
the program. The instrument can output arbitrary and sequenced
waveforms only after waveform data has been downloaded into its
memory.
Use the following command to select the output function type:
FUNCtion:MODE {FIXed | USER | SEQuence}
When “FIXed” is selected, the generator outputs the standard
waveform currently selected by the FUNC:SHAP command. When
“USER” is selected, the generator outputs the arbitrary waveform
currently selected by the TRAC:SEL command. When “SEQuence” is
selected, the generator outputs the sequence that is programmed
using the SEQ:DEF command. The query:
FUNCtion:MODE?
Queries the output function type and returns FIX, USER or SEQ.
Figure 3.6 Below shows how the Instrument control panel in WaveWiz
can be used select the waveform type..
Configuring the Instrument
27