4-19
Frequency Tuning Step Size
The Tuning Step Size can be adjusted using the UP and DOWN
arrow keys or the numeric entry keys (0.01 to 100.00 MHz in 10 kHz
steps).
Fundamental Hum Frequency
The fundamental hum frequency to be measured can be selected
using the
UP and DOWN arrow keys (60 Hz, 50 Hz, Auto). When
Auto is selected the unit will automatically switch to 50 Hz on PAL
type plans and 60 Hz for NTSC plans.
Scan Rate
Two scan rates are available in the Scan mode, normal and fast. The
scan rate is selected using the
UP and DOWN arrow keys. The Fast
scan feature allows for rapid scan displays, while sacrificing accuracy.
Normal scan rate is slower, but more accurate.
Scan Audio Carriers
If desired, the audio carriers can be omitted resulting in a faster scan.
This feature is selected using the
UP and DOWN arrow keys (yes,
no).
C/N Calibration
To increase the accuracy of the C/N measurements, a noise floor
calibration is performed. This allows the user to characterize the
noise floor of the unit. To perform the calibration, highlight the
selection and press the ENTER key. The unit will provide a prompt
to ensure that no cable is connected to the RF input. Once verified,
press the OK soft key to perform the calibration. The unit will
display the system noise level when the calibration is complete.
4.3.3 How To Setup Your Channel Plan
The procedure explained in Section 4.2.2 should be performed when
first using the Model 3HRV.